Patient Monitor User Manual Product Specification
- 253 -
Alarm Type SYS, DIA, MAP
Cuff pressure Measuring Range 0 mmHg to 300 mmHg
Pressure Resolution 1 mmHg
Measuring Accuracy
☆ Maximum Mean
Error
±5 mmHg
☆ Maximum Standard
Deviation
8 mmHg
Adult/Pediatric Neonate
Normal
Condition
Single Fault
Condition
Normal
Condition
Single fault
Condition
Maximum Cuff
Pressure
300 mmHg 330 mmHg 150 mmHg 165 mmHg
Maximum Measuring
Period
Less than 160 s Less than 180 s Less than 80 s Less than 90
s
Pre-inflation Pressure
Adult/ Pediatric Mode Default: 180 mmHg
Range: 120/140/150/160/180/200/220/240/260/280 mmHg
Neonatal Mode Default: 120 mmHg
Range: 80/100/120/140 mmHg
Dual Independent Channel Overpressure Protection
Adult/Pediatric < 300 mmHg
Neonatal < 150 mmHg
SunTech Module
Not applicable to iM50 or M50.
Method Oscillometric
Mode Manual, Auto, Continuous, Sequence
Measuring Interval in AUTO Mode
(unit: minute)
1/2/3/4/5/10/15/30/60/90/120/180/240/360/480 and
User Define
☆Measuring Parameter
SYS, DIA, MAP, PR
Pressure Unit kPa, mmHg, cmH
2
O
☆Measuring Range