Patient Monitor User Manual Product Specification
- 254 -
☆Adult Mode
SYS: 40 mmHg ~ 260 mmHg
DIA: 20 mmHg ~ 200 mmHg
MAP: 26 mmHg– 220 mmHg
☆Pediatric Mode
SYS: 40 mmHg– 230 mmHg
DIA: 20 mmHg– 160 mmHg
MAP: 26 mmHg– 183 mmHg
☆Neonatal Mode
SYS: 40 mmHg – 130 mmHg
DIA: 20 mmHg– 100 mmHg
MAP: 26 mmHg – 110 mmHg
☆Alarm Type
SYS, DIA, MAP
Pressure Resolution 1 mmHg
☆Maximum mean error
±5 mmHg
☆Maximum standard deviation
8 mmHg
Maximum measuring period
Adult 130 s
Pediatric 90 s
Neonate 75 s
Overpressure protection
Adult/Pediatric < 300 mmHg
Neonate < 150 mmHg
Pre-inflation Pressure
Adult Mode Default: 160 mmHg
Range:120/140/150/160/180/200/220/240/260/280 mmHg
Pediatric Mode Default: 140 mmHg
Range:
80/100/120/140/150/160/180/200/220/250 mmHg
Neonatal Mode Default: 90 mmHg
Range: 60/70/80/90/100/120/140 mmHg