Patient Monitor User Manual Product Specification
- 261 -
Halothane 4% None
Enflurane 5% None
Isoflurane 5% None
Sevoflurane 5% None
Xenon Not applicable Not applicable
Hehelium Not applicable Not applicable
Metered dose inhaler propellants Not applicable Not applicable
Desflurane 15% None
Ethanol 0.1% None
Isopropanol 0.1% None
Acetone 0.1% None
Methane 1% None
Respironics Sidestram and Mainstram Modules
Applicable Patient Type Adult, pediatric and neonatal patients
Technique Infra-red Absorption Technique
Measure Parameters EtCO
2
, FiCO
2
, AwRR
Unit mmHg, %, Kpa
Measuring Range
EtCO
2
0 mmHg to 150 mmHg
FiCO
2
3 mmHg to 50 mmHg
AwRR
0 rpm to 150 rpm (Mainstream)
2 rpm to 150 rpm (Sidestream)
Resolution EtCO
2
1 mmHg
FiCO
2
1 mmHg
AwRR 1 rpm
EtCO
2
Accuracy
± 2 mmHg, 0 mmHg to 40 mmHg
± 5 % of reading, 41 mmHg to 70 mmHg
± 8 % of reading, 71 mmHg to 100 mmHg
± 10 % of reading, 101 mmHg to 150 mmHg
± 12% of reading, RR is over 80 rpm (sidestream)
There will be no degradation in performance due to