Patient Monitor User Manual Product Specification
- 262 -
Respiration Rate. (mainstream)
AwRR Accuracy
± 1 rpm
Operation Mode Measure, standby
Sample Gas Flowrate
(sidestream)
(50 ±10) ml/min
O
2
Compensation
Range 0% to 100%
Resolution 1%
Default 16%
Barometric Pressure
Compensation
User setup
Anesthetic Gas Compensation
Range 0% to 20%
Resolution 0.1%
Default 0.0%
Balance Gas Compensation Room air, N
2
O, helium
Stability
Short Term Drift Drift over 4 hours < 0.8 mmHg
Long Term Drift 120 hours
Zero Calibration Support
Alarm Type
EtCO
2
, FiCO
2
, AwRR
Apnea Alarm Delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
Data Sample Rate 100 Hz
CO
2
Rise Time/Response Time
(mainstream)
Less than 60 ms
Sensor Response Time
(sidestream)
< 3 seconds, including transport time and rise time
Interfering Gas and Vapor Effects on EtCO
2
Measurement Values:
Gas or Vapor Gas Level (%) Quantitative Effect/Comments
Nitrous oxide
Halothane
60
4
Dry and Saturated Gas
(0 – 40) mmHg: ± 1 mmHg additional error

Table of Contents