Patient Monitor User Manual Product Specification
- 263 -
Enflurane
Isoflurane
Sevoflurane
Xenon
Helium
Desflurane
5
5
5
80
50
15
(41 – 70) mmHg: ± 2.5% additional erro
r
(71 – 100) mmHg: ± 4% additional error
(101 – 150) mmHg: ± 5% additional error
*Additional worst case error when compensation
for P
B
, O
2
, N
2
O, anesthetic agents, or helium is
correctly selected for the actual fractional gas
constituents present.
Desflurane:
The presence of desflurane in the exhaled breath at
concentrations greater than 5% will positively bias
Carbon Dioxide values by up to an additional 3
mmHg at 38 mmHg.
Xenon:
The presence of Xenon in the exhaled breath will
negatively bias Carbon Dioxide values by up to an
additional 5 mmHg at 38 mmHg.
Barometric Pressure on EtCO
2
Measurement Values:
Quantitative Effect
Ambient Barometric, Operational
(0 – 40) mmHg: ± 1 mmHg additional error
(41 – 70) mmHg: ± 2.5% additional error
(71 – 100) mmHg: ± 4% additional error
(101 – 150) mmHg: ± 5% additional error
*Additional worst case error when compensation for P
B
, O
2
, N
2
O, anesthetic agents, or helium
is correctly selected for the actual fractional gas constituents present.
NOTE:
Respiration Rate accuracy was verified by using a solenoid test setup to deliver a square
wave of known CO
2
concentration to the device. 5% and 10% CO
2
concentrations were
used. Respiration rate was varied over the range of the device. Pass/Fail criteria was
comparison of the respiratory rate output from the sensor to the frequency of the square
wave.
Masimo Sidestream Module
Dimensions 91 mm × 60 mm × 37 mm (3.6 × 2.4 ×1.4), exclduing
cable, tubing and Nomoline.
Weight 98 g, exclduing cable, tubing and Nomoline.
According to the degree of
protection against harmful
ingress of water
IP34