Patient Monitor User Manual Product Specification
- 276 -
N
2
O
4
60 vol%
1
1
1
HAL
4
4 vol%
1
1
1
ENF, ISO, SEV
4
5 vol%
1
1
1
DES
4
15 vol%
1
1
1
Xe (Xenon)
4
80 vol% -10% of reading
3
1
1
He (Helium)
4
50 vol% -6% of reading
3
1
1
Metered Dose Inhaler
Propellants
4)
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol)
4
0.3 vol%
1
1
1
C
3
H
7
OH
(Isopropanol)
4)
0.5 vol%
1
1
1
CH
3
COCH
3
(Acetone)
4
1 vol%
1
1
1
CH
4
(Methane)
4
3 vol%
1
1
1
CO
(Carbon monoxide)
5)
1 vol%
1
1
1
NO
(Nitrogen monoxide)
5)
0.02 vol%
1
1
1
O
2
5
100 vol%
2
1
1
Note 1: Negligible interference, effect included in the specification “Accuracy, all conditions”
above.
Note 2: Negligible interference with N
2
O / O
2
concentrations correctly set, effect included in the
specification “Accuracy, all conditions” above.
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically decreases the
CO
2
readings by 6%. This means that if measuring on a mixture containing 5.0vol% CO
2
and
50vol% Helium, the actual measured CO
2
concentration will typically be (1-0.06) * 5.0vol% =
4.7vol% CO
2
.
Note 4: According to the EN ISO 80601-2-55:2011 standard.
Note 5: In addition to the EN ISO 80601-2-55:2011 standard.
Dräger Minimodule
Method Sidestream gas measurement
Infrared measurement: CO
2
, N
2
O, anesthetic agents
Paramagnetic measurement: O
2
Barometric Pressure
Compensation
Automated compensation