Patient Monitor User Manual Product Specification
- 277 -
Gas Sampling Rate 200 mL/min ±20 mL/min
Maximum time until water trap
requires draining
41 hrs (sample gas under BTPS conditions, ambient air
23 °C)
Total System Response Time < 3 s
Drift Compensation (zeroing) Automated cyclical zeroing,
once per day (in error-free operation)
Zeroing Duration < 20 s
Cross Sensitivity None concerning alcohol (< 3000 ppm blood conc.),
acetone (< 1000 ppm), methane, water vapor, NO, and CO
O
2
Range
(0 to 100) Vol%
Accuracy
1
±(2.5 Vol% + 2.5 % rel.)
Rise Time (t10…90)
4
< 500 ms
Time to Specified Accuracy
3
< 450 s
CO
2
Range
(0 to 13.6) Vol%
Accuracy
1
±(0.43 Vol% + 8% rel.)
Rise Time (t10…90)
4
< 350 ms
Time to availability
2
< 60 s
Time to Specified Accuracy
3
< 450 s
N
2
O
Range
(0 to 100) Vol%
Accuracy
1
±(2 Vol% + 8% rel.)
Rise Time (t10…90)
4
< 350 ms
Time to Specified Accuracy
3
< 450 s
Anesthetic Gases Range
Halothane
(0 to 8.5) Vol%
Isoflurane
(0 to 8.5) Vol%
Enflurane
(0 to 10) Vol%
Sevoflurane
(0 to 10) Vol%
Desflurane
(0 to 20) Vol%