Patient Monitor User Manual Alarm Information
- 76 -
PR(NIBP) EDAN 40~240
Omron Adult/pediatric: 40~200
Neonate: 40~240
SunTech 30~220
PR (IBP) EDAN 30~300
NIBP alarm limits are listed as follows: unit (mmHg)
Patient Type
Adjustable Range
EDAN Omron SunTech
ADU SYS
DIA
MAP
25~290
10~250
15~260
60~250
40~200
45~235
40~260
20~200
26~220
PED SYS
DIA
MAP
25~240
10~200
15~215
60~250
40~200
45~235
40~230
20~160
26~183
NEO SYS
DIA
MAP
25~140
10~115
15~125
40~120
20~90
30~100
40~130
20~100
26~110
TEMP alarm limits are listed as follows:
Adjustable Range
T1
0 °C (32 ºF)~50 °C (122 ºF)
T2
0 °C (32 ºF)~50 °C (122 ºF)
TD
High limit: 0.1 °C (32.18 ºF)~50 °C (122 ºF)
IBP alarm limits are listed as follows: unit (mmHg)
Adjustable Range
Art 0~300
RAP/LAP/CVP/ICP -10~40
PA -6~120
P1/P2 -50~300
CO
2
alarm limits are listed as follows:
Adjustable Range
EtCO
2
0 mmHg~150 mmHg
FiCO
2
High limit: 3 mmHg~50 mmHg