Patient Monitor User Manual Alarm Information
- 77 -
AwRR Sidestream: 2 rpm ~150 rpm (EDAN, Respironics)
0 rpm ~150 rpm (Masimo)
Mainstream: 0 rpm~150 rpm
C.O. alarm limits are listed as follows:
Adjustable Range
TB
23 °C (73.4 ºF)~43 °C (109.4 ºF)
Quick Temp alarm limits are listed as follows:
Patient Type Adjustable Range
Quick Temp
ADU
35.5 °C (95.9 ºF)~42 °C (107.6 ºF)
PED
35.5 °C (95.9 ºF)~42 °C (107.6 ºF)
AG alarm limits are listed as follows:
EDAN Module
Adjustable Range
EtCO
2/
FiCO
2
0.0%~15.0%
EtO
2/
EtN
2
O/FiN
2
O 0.0%~100.0%
FiO
2
18.0%~100.0%
EtHAL/EtENF/EtISO/FiHAL/FiENF/FiISO 0.0%~8.0%
EtSEV/FiSEV/ 0.0%~10.0%
EtDES/FiDES 0.0%~22.0%
AwRR 0 rpm~150 rpm
Apnea Time 20 s~40 s
Masimo Module
Adjustable Range
FiCO
2
0.1%~25.0%
EtCO
2
0.0%~25.0%
FiO
2
18.0%~100.0%
EtO
2
0.0%~100.0%
FiN
2
O/EtN
2
O 0.0%~82.0%
EtDes/FiDes 0%~18%
EtIso/FiIso/EtHal/FiHal/EtEnf/FiEnf 0%~5.0%
EtSev/FiSev 0%~8%
AwRR 0 rpm~150 rpm
Apnea Time 20 s~40 s