VE-H100B Veterinary Pulse Oximeter User Manual
- 91 -
relative
humidity should
be at least 30%.
Electrical Fast
Transient/Burst
IEC/EN61000-4-4
±2kV for
power
supply lines
±2kV for
power supply
lines
Mains power
quality should
be that of a
typical
commercial or
hospital
environment.
Surge
IEC/EN61000-4-5
±1 kV line
to line
±2 kV line
to ground
±1 kV line
to line
±2 kV line
to ground
Mains power
quality should
be that of a
typical
commercial or
hospital
environment.
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
<5%UT(>95
% dip in
UT)for 0.5
cycle
<5% UT
(>95% dip in
UT) for 0.5
cycle
Mains power
quality should
be that of a
typical
commercial or
理邦保密文件