VE-H100B Veterinary Pulse Oximeter User Manual
- 45 -
Alarm, there will be↑or↓icon displayed on screen, and the SpO
2
or PR character will flash.
If Alarm system is set to OFF, all audio alarms and visual alarms
are turned off.
WARNING
When the Alarm system is off, the oximeter will not give an
alarm prompt. In order to avoid endangering the patient’s life,
the user should use this function cautiously.
3.3.10 SpO
2
Alarm Setup
The user can choose SpO
2
Hi Alarm and SpO
2
Lo Alarm in
menu to adjust SpO
2
alarm limit. Press the Up Arrow button or
Down Arrow button to increase or decrease alarm limit.
By default, SpO
2
Hi Alarm and SpO
2
Lo Alarm are 100 and 85
respectively.
Set the SpO
2
alarm limits as follows:
Choose SpO
2
Hi Alarm in the menu, press the
Function button to enter setup. The SpO
2
Hi Alarm
box will change from real line box to broken line box.
The adjustable range for upper limit of SpO
2
is from
“1 + the lower limit of SpO
2
” to 100. If the value of
SpO
2
Hi Alarm is set to less than 85, it will restore to
理邦保密文件
EDAN CONFIDENTIAL