VE-H100B Veterinary Pulse Oximeter User Manual
- 86 -
Appendix II EMC Information
Refer to the following tables for specific information regarding
this device’s compliance to IEC/EN 60601-1-2.
A2.1 Electromagnetic Emissions
For pulse oximeter and charger stand:
Guidance and manufacturer’s declaration – electromagnetic
emissions
VE-H100B and charger stand are intended for use in the
electromagnetic environment specified below. The customer or
the user of the VE-H100B and charger stand should assure that
they are used in such an environment.
Emissions test
Compliance
Electromagnetic
environment -guidance
RF emissions
CISPR11
Group 1
VE-H100B and charger
stand uses RF energy only
for its internal function.
Therefore, its RF
emissions are very low and
are not likely to cause any
interference in nearby
electronic equipment.