VE-H100B Veterinary Pulse Oximeter User Manual
- 46 -
default value after the oximeter is turned on again.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
Choose SpO
2
Lo Alarm in the menu, press the
Function button to set up it. The SpO
2
Lo Alarm box
will change from real line box to broken line box. The
adjustable range for the lower limit of SpO
2
Alarm is
from 0 to the upper limit of SpO
2
Alarm - 1. If the
value of SpO
2
Lo Alarm is set to less than 85, it will
restore to 85 after the oximeter is turned on again.
Press the Up Arrow or Down Arrow button to
increase or decrease values.
SpO
2
Hi Alarm is always higher than SpO
2
Lo Alarm
by at least 1%.
Press the Function button, confirm the alarm range
setup.
Press the On/Off button to exit the menu, and return to
measurement state.
3.3.11 PR Alarm Setup
The user can use PR Hi Alarm and PR Lo Alarm in menu to
adjust pulse rate alarm limits.
By default, PR Hi Alarm and PR Lo Alarm are 170 and 40
respectively.