VE-H100B Veterinary Pulse Oximeter User Manual
- 98 -
communications equipment.
Rated
maximum
output power
of transmitter
(W)
Separation distance according to
frequency of transmitter (m)
150 kHz to
80 MHz
Pd 2.1
80 MHz to
800 MHz
Pd 2.1
800 MHz to 2.5
GHz
Pd 3.2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed
above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation
distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.