EasyManua.ls Logo

Edwards iF1800 - Page 116

Edwards iF1800
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A533-23-880 Issue C
Page 106 © Edwards Limited 2009. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
maintenance
WARNING 5613 You have configured the iF to have an exhaust
temperature sensor fitted, but the sensor is not
fitted, or is disconnected or has failed. Check and
rectify as necessary.
WARNING 5709
DP TEMP LOW
DP TEMP LOW
SEE MANUAL
The operating temperature of the FDP pump is too
low:.
WARNING 5711
DP TEMP HIGH
DP TEMP HIGH
SEE MANUAL
The operating temperature of the FDP pump is too
high: there may be process debris in the pump or the
pump may be damaged.
WARNING 5713 The FDP pump-body temperature sensor is not fitted,
or is disconnected or has failed. Check and rectify as
necessary.
WARNING 6001
WATER FLOW LOW
CHECK COOLING
WATER SUPPLY
The flow of cooling-water through the iF system is too
low. Refer to the recommended actions for error
number 5411.
WARNING 11101 There is a fault in the NIM (Network Interface
Module): refer to the instruction manual supplied
with the NIM.
WARNING 12101 There is a fault in the Tool Interface Card in the iF
Interface Module. Refer to the instruction manual
supplied with the iF Interface Module.
WARNING 15101 There is a fault in the Auxiliary Interface Card in the
iF Interface Module.
Refer to the instruction manual supplied with the iF
Interface Module.
WARNING 15201 The gate valve connected to the iF Interface Module
has failed to close.
WARNING 15301 The gate valve connected to the iF Interface Module
has failed to open.
WARNING 17601 SEE MANUAL The inverter on an iF1800 system has experienced a
transient power supply problem, or there is a
recoverable fault.
WARNING 17613 SEE MANUAL There is a fault with the RS485 data link between the
inverter drive and the control module. Check that the
lead is connected.
WARNING 17701 SEE MANUAL The HMB pump stator temperature is > 125
o
C, and
the HMB pump speed has been reduced to 1200 r min
-
1
. The warning condition will be removed and the
HMB pump speed will increase to 6000 r min
-1
again
when the pump stator temperature has fallen to 90
o
C.
Table 19 - Warning messages
Warning message Advisory message Meaning/recommended action

Table of Contents

Related product manuals