A540-55-880 Issue D
Page 22 © Edwards Limited 2015. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Installation
Section 2.5.
Note: Pump rating information can be found on the label on the rear of the pump.
If connecting the electrical supply to the system through an ELCB (or RCD depending on territory) it must be suitable
for protection of equipment with a d.c. component in the fault current, and suitable for short-duration, switch-on
surges, and for high-leakage current (for example, type B, according to EN50178).
The secondary protective earth (ground) is required in case of failure of the primary earth and because pump filters
can cause high earth leakage currents, refer to Table 7.
3.7.1 Mains supply cable connection
Use the following procedure (refer to Figure 9) to connect the electrical supply to the dry pump system using the
connector supplied with the pump. If using a different connector to that supplied, follow the manufacturer's
instructions. When making the electrical supply cable, ensure that the earth (ground) conductor is longer than the
phase conductors. This will ensure that if the cable is accidentally dragged and the strain relief bush on the electrical
supply connector mating half fails, the earth (ground) conductor will be the last conductor to be pulled from the
connector.
If further information on connecting the electrical supply is required, contact Edwards for advice.
1. Screw the cable gland onto the connector hood.
2. Pass a suitable cable through the cable gland and hood. The cross sectional area of the cable wires should be 6
mm
2
. Cables must be bare ended without ferrules to ensure correct clamping in the connector block.
3. Fit the coding pins to the connector block according to the branch supply voltage as shown in Figure 9. 200 V
supply is Low Volts, 400 V supply is High Volts.
Figure 9 - High and Low voltage coding pin arrangement
4. Before starting the assembly, use the hex (Allen) key as shown in Figure 11 to ensure that the axial cone is
screwed fully anticlockwise to completely open the contact chamber.
WARNING
The mains connector is not approved for connection and disconnection under load.