Descripon
Units
Rang
▪ for 380 - 460 V a.c. systems
mA < 10
* Typical earth leakage values were measured at steady-state condions.
Not
e that higher leakage currents can occur:
i) in transient condions such as power on or pump acceleraon
ii) with abnormal supply conguraons such as a missing or earthed phase or
unbalanced supply voltages.
Contact us for more informaon about conguraon requirements for earth leakage
reducon.
Table 9
Electrical connecons
Descripon Mang connector descripon / external
supply rang
Internal supply rang
Mains c
onnecon
R
efer to Installaon on page 25
for wiring diagram
Refer to Table: Pump electrical data for the
mains connector installed to each variant.
Insert type is one of:
Harng Han 40A, axial module, part num-
ber: 09 14 002 2701, 2.5 –
8 mm2 ne stran-
ded wire, Hood part number: 09 30 006
0442 or
Insert type is : Weidmüller 1789980000, 2.5
– 8 mm2 ne stranded wire, Hood part
number : 1652680000 (VDE 0295 class 5, re-
fer to Table: Wire assembly according to
VDE 0295), 8.9 mm maximum insulaon di-
ameter or
Han K 4/4-F nger safe 09 38 008 2703, 6-16
mm
2
ne stranded wire (VDE 0295 class 5,
refer to Table: Wire assembly according to
VDE 0295), 8.9 mm maximum insulaon di-
ameter
PDT interface (front) XLR type 5-way plug 24 V d.c. 0.2 A
System interface (rear) XLR type 5-way plug
24 V d.c. 0.75 A
*
Ethernet interface Standard RJ45 type or Neutrik® EtherCon®
RJ45
(IEEE802.3i 10 Base T
E
therne
t)
USB ash memory sck/drive USB Type A ash sck only (no cable con-
necon allowed)
5 V d.c. 100 mA
EMS interface XLR type 6-way plug
External controlled stop buon
Pin 1 supply, Pin 2 - r
eturn
Internal controlled stop buon§
24 V d.c. 100 mA
10/2020 - ©Edwards Limited
Page 20M56635880_D
M56635880_D - Technical data