Remedy Refer to Unplanned shut down and alarms on pag
e 80 to determine the exhaust
pressure that caused the warning.
Check that all valves in the exhaust line are open and consider whether process
debris or condensaon are likely.
Fault Warning 39.13 - Exh Press Fault
Cause Sensor is faulty or disconnected
Remedy Contact us.
Fault Warning 54.11 - MB Mtr Temp High
Cause The temperature of the booster pump is too high.
Remedy Check that cooling water is connected, switched on and is to specicaon giv
en in
Cooling water data on page 21.
Fault Warning 54.13 - Sensor missing
Cause The booster temperature sensor has possibly become disconnected or failed.
Remedy Check that the sensor is installed and is correctly connected. Cooling wat
er dat
a on
page 21
Fault Warning 54.13 - Sensor Fault
Cause Sensor is faulty (short circuit).
Remedy Check that the sensor is installed and is correctly connected.
Fault Warning 55.11 - DP Temp High
Cause Pump temperature high
Remedy Check that cooling water is connected, switched on and is to specicaon given in
Cooling water data on page 21.
Fault Warning 55.13 - Sensor missing
Cause The dry pump temperature sensor is possibly disconnected or failed.
Remedy Check that the sensor is installed and is correctly connected.
Fault Warning 55.13 - Sensor Fault
Cause Sensor is faulty (short circuit).
Remedy Check that the sensor is installed and is correctly connected.
10/2020 - ©Edwards Limited
Page 71M56635880_D
M56635880_D - Fault nding