Some of these warnings apply to accessories and can only be encountered if the
applicable ac
cessory is installed.
Fault Warning 1.01 - Power interrupt
Cause There has been a brownout of the electrical supply to the pump lasng more
than 1 sec
ond.
Remedy Check the electrical supply.
Fault Warning 31.13 - Gas missing
Cause The gas module is not communic
ang with the pump controller.
Remedy Contact us.
Fault Warning 31.01 - SYS CONFIG FAULT
Cause The wrong type of gas module has been installed.
Remedy Contact us.
Fault Warning 35.09 - N2Purge Low
Cause The nitrogen supply ow
rate is too low.
▪ The supply is possibly not connected or set to on.
▪ The supply line or lter is possibly blocked.
Remedy Check the nitrogen supply is connected and set to on and that pipelines and lters
are not blocked.
Fault Warning 35.11 - N2 Purge High
Cause The Nitrogen supply ow rate is too high.
▪ The gas module and PDT sengs do not mat
ch.
Remedy Check the gas module and PDT sengs, re
fer to Gas module conguraon on page
64.
Fault Warning 35.13 - Gas Sensor Fault
Cause The Nitrogen supply o
w
rate is too high.
▪ The gas module is possibly faulty or disconnected
Remedy Contact us.
Fault Warning 39.11 - Exh Press High
Cause The pressure in the exhaust pipeline is too high.
▪ A valve in the pipeline is possibly shut.
▪ There is possibly process debris or c
ondensaon in the e
xhaust pipeline.
▪ You can have too many pumping systems connected to the exhaust pipeline.
10/2020 - ©Edwards Limited
Page 70M56635880_D
M56635880_D - Fault nding