Fault Event 252.63 - Chassis short
Cause Wiring fault on pump
Remedy Contact us.
Fault Event 252.65 - Chassis short
Cause Wiring fault on pump
Remedy Contact us.
Fault Event 253.63 - Accessory oload
Cause Faulty equipment has just been plugged in
Remedy
1. Examine the equipment recently connected to the pump. Make sure that is
ins
t
alled and correctly connected.
2. If the warning does not disappear power cycle the pump.
3. If the warning re-appears aer power cycling, contact supplier.
Fault Event 254.63 - Bus oload
Cause Faulty equipment has just been plugged in
Remedy
1. Examine the equipment recently connected to the pump. Make sure that is
inst
alled and correctly connected.
2. If the warning does not disappear power cycle the pump.
3. If the warning re-appears aer power cycling, contact us.
Fault Event 255.63 - Accessory oload
Cause Faulty equipment has just been plugged in
Remedy
1. Examine the equipment recently connected to the pump. Make sure that is
installed and correctly connected.
2. If the warning does not disappear power cycle the pump.
3. If the warning re-appears aer power cycling, contact us.
Fault Event 256.63 - 24 V oload
Cause Faulty equipment has just been plugged in
Remedy
1. Examine the equipment recently connected to the pump. Make sure that is
ins
t
alled and correctly connected.
2. If the warning does not disappear power cycle the pump.
3. If the warning re-appears aer power cycling, contact us.
Fault Event 257.63 - Front Port oload
Cause Faulty equipment has just been plugged in
10/2020 - ©Edwards Limited
Page 82M56635880_D
M56635880_D - Fault nding