7 Operation
16 / 31
7.2 Heating
Press to heat the heating needle.
The indicator LED lights up during heating.
WARNING
Do not place the heated heating needle in the root
canal for more than 4 seconds to prevent thermal
injury to the patient.
Only the end of the heating needle (about 4-5mm)
can be heated. Use this area to cut the gutta percha.
1
2 3
16 6
2
1
The “Keep-heat Time” will display on the screen.
When the set time is reached, the heating process will
stop.
2
Heating indication
3
Real time heating temperature
2
1
66
Release , the heating needle will cool down.
1
Cooling indication
2
Real time cooling temperature
When the set time of “CoolingDisplay” is reached, the
screen will switch to the standby interface.
Tip error
If the heating needle is not installed correctly, or the
heating needle is broken, the “Tip error” will appear.
Heating area