11 EMC Tables
25 / 31
11.EMC Tables
Guidance and manufacturer’s declaration – electromagnetic emissions
The Fast-Pack pro is intended for use in the electromagnetic environment specified below.
The customer or the user of the Fast-Pack pro should assure that it is used in such an
Emissions test
Complian
Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1
The Fast-Pack pro uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The Fast-Pack pro is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
Harmonic emissions
Class A
Voltage fluctuations/flicker
emissions
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The Fast-Pack pro is intended for use in the electromagnetic environment specified below.
The customer or the user of the Fast-Pack pro should assure that it is used in such an
Immunity test IEC 60601 test
Compliance
Electromagnetic
Electrostatic
discharge
(ESD) IEC
61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4
kV, +/- 8 kV, +/-
15 kV air
+/- 8 kV contact
+/- 2 kV, +/- 4
kV, +/- 8 kV, +/-
15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.