11.EMC Tables
Guidance and manufacturer’s declaration electromagnetic emissions
The Fast-Pack pro is intended for use in the electromagnetic environment specified below.
The customer or the user of the Fast-Pack pro should assure that it is used in such an
environment.
Emissions test
Complian
ce
Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1
The Fast-Pack pro uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The Fast-Pack pro is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration electromagnetic immunity
The Fast-Pack pro is intended for use in the electromagnetic environment specified below.
The customer or the user of the Fast-Pack pro should assure that it is used in such an
environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
+/- 8 kV contact
+/- 2 kV, +/- 4
kV, +/- 8 kV, +/-
15 kV air
+/- 8 kV contact
+/- 2 kV, +/- 4
kV, +/- 8 kV, +/-
15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.