at point of use
contaminations from the components, and
put them in container for transportation.
Prepare the components directly after
treatment.
Do not submerge the components or wipe
them with any of the following functional
water (acidic electrolyzed water, strong
alkaline solution, or ozone water), medical
agents (glutaral, etc.), or any other special
types of water or commercial cleaning liquids.
Such liquids may result in metal corrosion
and adhesion of the residual medical agents
to the components.
2
Preparation
before cleaning
Remove and disconnect the components
(Tips, Wrench, and Insulating sleeve) before
cleaning. Refer to "Chapter 4- Setting up the
Ultra X" of this manual for disassembly
instructions.
Observe suitable personal protective
measures.
The following Step 2 to Step 4 are operated in a washer-disinfector:
WARNING
Use only approved washer-disinfectors according to EN ISO 15883,
maintain and calibrate it regularly.
Follow instructions and observe concentrations given by the
manufacturer (see general recommendations).
Sufficient rinsing step should be available in purified water (max 10
germs/ml and max 0.25 endotoxin units/ml)
Avoid any contact between the tips and any instrument, kit, support or
container.
Make sure the components are dry before moving to the #5 step.
3
Cleaning:
Automated
Carefully put the components (Tips, Wrench,
and Insulating sleeve) into the washer-
disinfector and set the parameters as follows:
- Pre-cleaning: water temperature <30, 2