- Cleaning: water temperature 45℃, 5 min;
use an enzyme detergent solution (mild
and aldehyde free solution) which is
suitable to be used with washer-
disinfector, and use in accordance with
the IFU of the detergent solution
manufacturer;
- Rinsing: water temperature 45℃, 1 min
(rinsing twice).
Thermal disinfection at least 5 min at 90℃
/194°F, make sure A0 value≥3000.
Maintenance
and Inspection
Inspect components and sort out those with
defects. Dirty components must be cleaned
and disinfected again.
Pack each component in a separate steam-
sterilization pouch.
Check the validity period of pouch given
by the manufacturer to determine the
shelf life.
Use pouches which resist to a
temperature up to 141℃(286°F) and in
accordance with EN ISO 11607.