Installation
16
Fig. 6: Window Configuration Heater - Preheating
The value shows the chosen heating rate in °C/min. (The preheating function is
switched off by entering 0 in this place. Then, the heating occurs with maximum
speed after start of the UNI-software.)
3. Save your entry with Enter.
4. The preheating starts automatically with the next start of the software, if the
furnace temperature is lower than 100°C and the analyzer switch is at
position 2.
NOTICE
When the preheating function is activated (by any other value except 0) the
software automatically creates a shortcut at the startup menu of Windows. This
enables the start of the UNI-software with every start of the PC. The shortcut is
deleted automatically when the preheating is deactivated.
2.9.2 BIOS-configuration
In order to start the preheating before the work begins, the BIOS has to be
configured to start the PC automatically at the selected time. (The preheating starts
automatically by starting the UNI-software.)
1. Switch on your PC and enter the BIOS by pressing the corresponding key
(e.g. F2). The normal booting process is interrupted and the BIOS menu
appears.
2. Choose Power Management Setup
3. Activate "Resume by Alarm" (enabled).
4. Enter the required days and times when the heating up of the analyzer
should be started.
5. Save your entries at Save/Exit with Yes.
6. The booting process will be sustained.

Other manuals for ELTRA CS-580A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA CS-580A and is the answer not in the manual?

ELTRA CS-580A Specifications

General IconGeneral
BrandELTRA
ModelCS-580A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals