EM Test dito
Instruction Manual V 2.30 26 / 36
Maintenance and practical advice
General
dito is maintenance free.
Calibration and Verification
Factory calibration
Every EM TEST generator is entirely checked and calibrated as per International Standard regulations before delivery. A
calibration certificate is issued and delivered along with a list of the equipment used for the calibration proving the traceability
of the measuring equipment. All auxiliary equipment and accessories are checked to our internal manufacturer guidelines.
The calibration certificate and the certificate of compliance (if available) show the date of calibration.
The EM Test equipment are calibrated in the factory and marked with a calibration mark. The
used measuring instruments are traceable to the Swiss Federal Office of Metrology.
The calibration date is marked. The validity of the calibration is to the responsibility of the
user’s quality system. Neither the certificate of calibration nor the corresponding label mark
any due date for re-calibration.
Example: Calibration mark
Guideline to determine the calibration period of EM Test instrumentation
Our International Service Departments and our QA Manager are frequently asked about the calibration interval of EM TEST
equipment.
EM TEST doesn’t know each customer’s Quality Assurance Policy nor do we know how often the equipment is used and
what kind of tests are performed during the life cycle of a test equipment. Only the customer knows all the details and there-
fore the customer needs to specify the calibration interval for his test equipment.
In reply to all these questions we like to approach this issue as follows :
EM TEST make use of a solid state semiconductor switch technique to generate high voltage transients. A precious ad-
vantage of this technique is the absolute lack of periodical maintenance effort. In consequence thereof a useful calibration
period has to be defined based on two criteria :
- The first one is the customer’s Quality Assurance Policy. Any existent internal regulation has to be applied at highest
priority. In the absence of such internal regulation the utilization rate of the test equipment has to be taken into considera-
tion.
- Based on the experience and observation collected over the years EM TEST recommend a calibration interval of 1
year for frequently used equipment. A 2-years calibration interval is considered sufficient for rarely used test generators
in order to assure proper performance and compliance to the standard specifications.
Calibration of Accessories made by passive components only
Passive components do not change their technical specification during storage. Consequently the measured values and the
plots stay valid throughout the storage time. The date of shipment shall be considered as the date of calibration.
Periodically In-house verification
Please refer to the corresponding standard before carrying out a calibration or verification. The standard describes the pro-
cedure, the tolerances and the necessary auxiliary means. Suitable calibration adapters are needed. To compare the verifi-
cation results, EM Test suggests to refer to the wave shape and values of the original calibration certificate.