EM Test dito
Instruction Manual V 2.30 29 / 36
Example of a calibration certificate
DIN EN ISO/IEC 17025
CALIBRATION CERTIFICATE
EM Test (Switzerland) GmbH, Sternenhofstrasse 15, CH- 4153 Reinach, www.emtest.com
Certificate No
CE-1211112234-569911-Dito
Order No
KD-999999
Date of Order
06.03.2012
Operator
P. Fuchs
Date of Calibration
21.03.2012
Customer
Name of the customer.
City, Country
Object
ESD Generator
Type
Dito
Serial No
V1211112233
Inventory No
- - -
Firmware Version
2.31
Device
212023
Modules
150pF / 330
Serial No
1602
150pF / 330Ω RFCI
Serial No
1601
Manufacturer
EM Test (Switzerland) GmbH
Remarks
The instrument has been fully tested and calibrated. It complies with the following standards and with
the manufacturers specification.
IEC 61000-4 part 2. Ed.2. : 2008
ISO 10605 : 2008
Calibration Results
The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage
factor k = 2, which for a normal distribution corresponds to a coverage probability of approximately 95%.
The measurement uncertainty is not taken in account for determining compliance with the specifications.
Traceability
This calibration certificate documents the traceability to national standards, which realize the physical units of measurements (SI).
The measurements, the uncertainties with confidence probability and calibration methods are given on the following pages and are part of
the certificate. The document CMAM000001D is used as base for the accredited calibration procedure.
Calibration Intervals of the calibrated equipment
The equipment should be re-calibrated at appropriate intervals.
Calibration Intervals of EM Test Measuring equipment
According to our accredited management system all used measuring equipment of EM Test is calibrated in cycles of two years.
Copies of the Certificate
This calibration certificate shall not be reproduced except in full, without written approval of the laboratory EM Test.
Certificates of EM Test are not valid without signature of the head of the laboratory.
Date
Head of the Calibration Laboratory
Name
Dito This report must not be reproduced except in full Page 1 of 5