WARNING!
The USB port must only be used in a non-hazardous location.
The Ethernet port must only be used in a non-hazardous location.
Do not use the CSI 430 SpeedVue Sensor in a hazardous location.
The battery must only be charged and/or replaced in a non-hazardous
location.
If a unit shows any sign of damage, please return for repair.
If leaving the device unattended outdoors, it is recommended to store
the unit in a shaded area or with the LCD facing down.
The front touch screen must be protected from impact.
Outputs are intrinsically safe when implemented per drawing D25671 for
use in an ATEX/IECEx Zone 2 hazardous environments.
Outputs are intrinsically safe when implemented per drawing D25639 for
use in a Class I Division 2 hazardous environment.
Refer to Emerson Safety Addendum D25670 for complete information on
certifications and conditions of safe use in ATEX/IECEx Zone 2 locations. Only
ATEX/IECEx Zone 2 units will include this safety addendum in the package.
Notes
To comply with relevant safety certifications, the CSI 2140 labeled "ATEX/
IECEx Zone 2" does not have a keypad backlight.
CSI 430 SpeedVue Sensor may not be compatible with the CSI 2140
labeled "ATEX/IECEx Zone 2." The CSI 430 is not permitted in hazardous
areas; and it may not function with the ATEX-certified CSI 2140 even in a
safe area.
Quick Start Guide March 2016
16 MHM-97430 Rev 4

Other manuals for Emerson CSI 2140

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson CSI 2140 and is the answer not in the manual?

Emerson CSI 2140 Specifications

General IconGeneral
BrandEmerson
ModelCSI 2140
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals