Table 2-2: Maintenance (continued)
Daniel MeterLink utility Diagnostics Action(s)
External ground wiring • Transmitter Electronics
Enclosure ground lug
• Inspect ground lug wiring
and make sure the wiring is
tightly secured.
Conduit seals • Transmitter Electronics
Enclosure
• Inspect the conduit sealant
and follow your standard
operating procedure to
report tampering with the
conduit sealant.
• Your operating procedures
may require a certified
electrician and company
witness to reseal the
conduit.
Flanges • Inspect for leaks
• Inspect flange stabilizers
• Perform leak tests on
flanges
• Ensure flange stabilizers are
installed.
WARNING
CRUSHING HAZARD
Do not remove flange stabilizers.
Attempting to do so may allow the meter to roll, resulting in
serious injury or equipment damage.
A. Flange stabilizers
2.2.2
Unable to connect direct serial or external serial modem
If you are using Windows
®
7 or Windows
®
8 make sure that you do not have more than
one modem driver installed to the same COM port. Typically this will only be necessary if
you use one COM port to talk direct (serial communications) and use the same COM port
to connect to an external modem. This is an apparent limitation in Microsoft’s
®
Dial-up
Networking. If more than one modem driver is installed for a particular COM port, Dialup
Networking will always use the last driver installed regardless of what is selected. The only
work around is to only install one modem driver per COM port on the PC at a time. Refer to
Troubleshooting Maintenance and Troubleshooting manual
June 2019 P/N 3-9000-769
44 Gas Ultrasonic Flow Meters