Table 2-2: Maintenance (continued)
Daniel MeterLink utility Diagnostics Action(s)
MeterLink Logs/ Reports Menu • Power failure • If this was a known power
fail or restart of the meter
just acknowledge this alarm
on the Status Summary
page.
• If this was an unexpected
restart of the meter, verify
the integrity of the power
to the meter and make sure
that the voltage level is in
the range of 11-36 VDC at
the meter.
• Collect a complete Archive
log and contact your local
area Daniel service
representative
Daniel MeterLink Meter
Monitor (Summary) view
• Chord failure Daniel
MeterLink
• The meter is unable to
obtain measurement data
from a pair of transducers.
• The cause may be isolated
to one pair of transducers
or its cabling. Check the
transducer wiring for this
pair of transducers to make
sure connections are secure
and wired correctly.
• Verify the average gain of
this transducer pair is not
above 90dB. Read the value
from the Daniel MeterLink
Monitor Page or using AMS
under Service Tools → Path
performance.
• Remove the transducer and
clean the transducer face
( Transducer removal
procedure).
Security seals
• Endcap seals
• Endcaps latches
• Transmitter Electronics
Enclosure
• Base Enclosure
• Shroud seals
• Only authorized personnel
may remove security seals.
Follow your standard
operating procedure to
report seals that have been
tampered with or removed
and replace the seals per
instructions in External
power source connection
and fuse in the Installation
Manual (3-9000-765).
Maintenance and Troubleshooting manual Troubleshooting