EasyManua.ls Logo

Emerson Daniel 3410 Series - Page 58

Emerson Daniel 3410 Series
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.3.2 T-Slot transducer removal without extractor tool
This procedure requires venting the line pressure prior to removing the transducer(s).
Before removing and installing transducer(s), connect to the meter using Daniel MeterLink
and collect and save a Maintenance Log.
NOTICE
The following instructions are for transducer removal and installation without the use of
an extractor tool. It is recommended that one work on one transducer assembly at a time
to reduce the possibility of improper assembly with respect to transducer lengths and
location.
Transducers are always replaced in pairs.
Update the calibration parameters for every chord when replacing a pair of
transducers. Refer to Modifying the calibration parameters for T-Slot transducers.
Table 3-2: Tools and supplies for disassembly and assembly
Tool list, shop supplies and reference documents
3/8” drive ratchet
3/8” drive extension
3/8” drive 7/16” deep socket
3/8” drive 5/16” hex socket
3/8” drive 1/2” hex socket
3/8” to 1/2” driver adapter
1/2” drive ratchet
1/2” drive 1 1/4” Socket
1 1/4” combination wrench
1 1/8” combination wrench
1” combination wrench
3/16” flat Blade screw driver
1/16” Xcelite Hex driver
6 mm Allen wrench
N.A.S. Loctite Nickel Anti-Seize 16 oz. Brush Top P/N 77164
Molykote 111
Shop supplies
Shop rags
Adhesive stickers for the cable nuts Sharpie Fine Point Permanent Marker (Black) Series # 30000, to
label transducers:
A1, A2, B1, B2, C1, C2, D1 and D2 for SeniorSonic
meters
A1, A2, B1 and B2 for JuniorSonic
meters)
Liquid leak detection fluid (customer supplied item)
Meter repairs Maintenance and Troubleshooting manual
June 2019 P/N 3-9000-769
58 Gas Ultrasonic Flow Meters

Other manuals for Emerson Daniel 3410 Series

Related product manuals