Daniel 3410 Series Gas Ultrasonic Meter Operations Manual Miscellaneous equations
3-9000-777 Rev E January 2018
Chord “L” dimension calculation 143
B.2 Chord “L” dimension calculation
The chord “L” dimension is calculated from the meter housing length as well as the
transducer pair lengths, mount lengths, holder lengths, and stalk lengths as shown in
Equation B-5. The transducer lengths are etched on the transducers. Likewise, the lengths
of the mounts, stalk assemblies, and transducer holders are also etched on the individual
components. The length of the meter body is found on the original calibration sheet
supplied with the meter.
Equation B-5 Chord “L” Dimension
where
=
chord “L” dimension (in) (LA ... LD)
=
meter housing length (in)
=
transducer 1 mount length (in)
=
transducer 2 mount length (in)
=
transducer 1 length (in)
=
transducer 2 length (in)
=
transducer 1 stalk length (in)
=
transducer 2 stalk length (in)
=
transducer 1 holder length (in)
=
transducer 2 holder length (in)
L
chord
L
MeterHou gsin
L
Mount1
L
Mount2
++=
L
Xdcr1
L
Stalk1
L
Hldr1
L
Xdcr2
L
Stalk2
L
Hldr2
L
chord
L
MeterHou gsin
L
Mount1
L
Mount2
L
Xdcr1
L
Xdcr2
L
Stalk1
L
Stalk2
L
Hldr1
L
Hldr2