148 MECHANICAL/ELECTRICAL TROUBLESHOOTING
Troubleshooting comm., mech., and elec. Daniel 3410 Series Gas Ultrasonic Meter Operations Manual
January 2018 3-9000-777 Rev E
communicating with meter but all
chords display failures
Verify that the resistance of transducers is within Specification (1-2 Ω).
Check the Acquisition board.
Check the interconnect cables between the base enclosure and the transmitter
enclosure.
chord is not indicating
Check the resistance of the failed transducer.
If Chord A is not indicating, change the transducer cables from Chord D to chord A.
If Chord D then fails, the transducers are bad on Chord A.
The same test procedure can be accomplished by swapping Chords B and C if a
chord fails in either chord.
Note that the outside chord cables cannot be exchanged with inner
chord cables.
Table C-1 Mechanical/Electrical Troubleshooting