Note
Do not apply lubricant to the transducer or stalk O-rings.
NOTICE
Ensure that the transducers identified as belonging to end 1 are installed on end 1 of the
meter housing and those identified as belonging to end 2 are installed on end 2 of the
meter housing.
5. Replace the O-ring and Backup O-ring on the transducer holder. It is highly
recommended that the O-rings be replaced when the transducer is removed from
the holder or stalk. Make sure that the contoured side of the ring is facing away from
the mount. Lubricate with Molykote111 Silicone Grease or equivalent.
6. Apply a small amount of nickel anti-seize (N.A.S.) compound (P/N 2-9-9960-134) to
the outer threads of the transducer holder (see Section 3.2.2).
7. Carefully install the transducer holder assembly into the transducer mount. Make
sure the threads of the holder and mount are correctly aligned. Use a 1 1/4” socket
and screw the transducer assembly into the mount. Tighten to securely seat the
assembly in the mount. Do not over tighten (see Figure 3-3).
8. Install the keyed transformer module into the transducer holder (see Figure 3-3).
a. Apply a small amount of Molykote 111 to the transformer module O-ring.
b. Insert the keyed transformer module into the back end of the transducer holder.