For IN and OUT events, set Setpoint A and Setpoint B.
5. (Optional) Configure a discrete output to switch states in response to the event
status.
6. (Optional) Specify the action or actions that the transmitter will perform when the
event occurs.
With the display: OFF-LINE MAINT > OFF-LINE CONFG > IO > CH C > SET DI > DI ACT
With ProLink II: ProLink > Configuration > Discrete Input
With ProLink III: Device Tools > Configuration > I/O > Action Assignment
With the Field Communicator: Configure > Alert Setup > Discrete Events > Assign
Discrete Action
Options for Enhanced Event Action
Options for Enhanced Event ActionTable 6-11:
Action Label
Display ProLink II ProLink III Field Communicator
Standard
None (default)
NONE None None None
Start sensor zero
START ZERO Start Sensor Zero Start Sensor Zero Perform auto zero
Start/stop all totaliz-
ers
START STOP Start/Stop All Totalization Start/Stop All Totalization Start/stop totals
Reset mass total
RESET MASS Reset Mass Total Reset Mass Total Reset mass total
Reset volume total
RESET VOL Reset Volume Total Reset Volume Total Reset volume total
Reset gas standard
volume total
RESET GSVT Reset Gas Std Volume
Total
Reset Gas Std Volume
Total
Reset gas standard vol-
ume total
Reset all totals
RESET ALL Reset All Totals Reset All Totals Reset totals
Meter verification
Start meter verifica-
tion test
START VERFY Start Meter Verification Start Meter Verification
Not available
CAUTION!
Before assigning actions to an enhanced event or discrete input, check the status of the event
or the remote input device. If it is ON, all assigned actions will be performed when the new
configuration is implemented. If this is not acceptable, wait until an appropriate time to assign
actions to the event or discrete input.
Integrate the meter with the control system
102 Micro Motion
®
Model 1700 Transmitters with Analog Outputs