5.6.4 Configure Sensor Serial Number
Display
Not available
ProLink II ProLink > Configuration > Sensor > Sensor S/N
ProLink III Device Tools > Configuration > Informational Parameters > Sensor
Field Communicator Configure > Manual Setup > Info Parameters > Sensor Information > Sensor Serial Number
Overview
Sensor Serial Number lets you store the serial number of the sensor component of your
flowmeter in transmitter memory. This parameter is not used in processing and is not
required.
Procedure
1.
Obtain the sensor serial number from your sensor tag.
2. Enter the serial number in the Sensor Serial Number field.
5.6.5 Configure Sensor Material
Display
Not available
ProLink II ProLink > Configuration > Sensor > Sensor Matl
ProLink III Device Tools > Configuration > Informational Parameters > Sensor
Field Communicator Configure > Manual Setup > Info Parameters > Sensor Information > Tube Wetted Material
Overview
Sensor Material lets you store the type of material used for your sensor’s wetted parts in
transmitter memory. This parameter is not used in processing and is not required.
Procedure
1.
Obtain the material used for your sensor’s wetted parts from the documents
shipped with your sensor, or from a code in the sensor model number.
To interpret the model number, refer to the product data sheet for your sensor.
2. Set Sensor Material to the appropriate option.