EasyManua.ls Logo

Emerson Micro Motion 1700 - Milliamp Output Problems

Emerson Micro Motion 1700
328 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10.6 Milliamp output problems
Milliamp output problems and recommended actionsTable 10-6:
Problem Possible causes Recommended actions
No mA output • Wiring problem
• Circuit failure
• Channel not configured for desired output
• Check the power supply and power supply
wiring. See Section 10.9.
• Check the mA output wiring.
• Check the Fault Action settings. See
Section 10.19.
• Measure DC voltage across output termi-
nals to verify that the output is active.
• Contact Micro Motion.
Loop test failed • Power supply problem
• Wiring problem
• Circuit failure
• Channel not configured for desired output
• Incorrect internal/external power configu-
ration
• Check the power supply and power supply
wiring. See Section 10.9.
• Check the mA output wiring.
• Check the Fault Action settings. See
Section 10.19.
• Contact Micro Motion.
mA output below 4
mA
• Open in wiring
• Bad output circuit
• Process condition below LRV
• LRV and URV are not set correctly
• Fault condition if fault action is set to inter-
nal zero or downscale
• Bad mA receiving device
• Check your process conditions against the
values reported by the flowmeter.
• Verify that the receiving device, and the
wiring between the transmitter and the re-
ceiving device.
• Check the settings of Upper Range Value and
Lower Range Value. See Section 10.18.
• Check the Fault Action settings. See
Section 10.19.
Constant mA output • Incorrect process variable assigned to the
output
• Fault condition exists
• Non-zero HART address (mA output 1)
• Output is configured for loop test mode
• Zero calibration failure
• Verify the output variable assignments.
• View and resolve any existing alarm condi-
tions.
• Check the HART address and Loop Current
Mode. See Section 10.15.
• Check to see if a loop test is in process (the
output is fixed).
• Check HART burst mode configuration.
See Section 10.16.
• If related to a zero calibration failure, cycle
power to the meter and retry the zeroing
procedure.
Troubleshooting
184 Micro Motion
®
Model 1700 Transmitters with Analog Outputs

Table of Contents