EasyManua.ls Logo

Emerson Micro Motion 1700 - Page 145

Emerson Micro Motion 1700
328 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Smart Meter Verification flowchart: Running a test using the display
Running a Smart Meter Verification test using the displayFigure 9-2:
OUTPUTS
ARE YOU SURE/YES?
. . . . . . . . . . . . . . . x%
PASS VERFY ABORTED VERFYCAUTION VERFY
Fail
Abort
RERUN/YES?
Yes No
Correct condition
RUN VERFY
CONTINUE MEASR FAULT LAST VALUE
Select
Scroll Scroll
Scroll
Select
Scroll
RESULTS VIEW/YES?
Select
Scroll Scroll
Select Select
Select
Select
Select
SENSOR ABORT/YES?
SelectScroll
Abort Type
Scroll
Scroll Select
To Runcount
(see Results Read)
Test resultPass
EXIT
Scroll
EXIT
Scroll
To Enter Meter Verfy
Run a Smart Meter Verification test using ProLink II
1.
Choose Tools > Meter Verification > Run Meter Verification.
Measurement support
Configuration and Use Manual 137

Table of Contents