Manage scheduled test execution using the display
1. Navigate to the Smart Meter Verification menu.
Smart Meter Verification – Top-level menuFigure 9-5:
Scroll and Select simultaneously
for 4 seconds
ENTER METER VERFY
Scroll
RUN VERFY RESULTS READ SCHEDULE VERFY
Select
EXIT
Scroll Scroll Scroll
Scroll Select
Select Select Select
2. Scroll to Schedule Verfy and press Select.
3.
To schedule a single test or the first test in recurring execution:
a. Scroll to Set Next and press Select.
b. Enter the number of hours that the transmitter will wait before beginning the
test.
4. To schedule recurring execution:
a. Scroll to Set Recur and press Select.
b. Enter the number of hours that will elapse between tests.
5. To disable scheduled execution:
To disable execution of a single scheduled test, set Set Next to 0.
To disable recurring execution, set Set Recur to 0.
To disable all scheduled execution, choose Turn Off Sched when you enter the
Smart Meter Verification menu.