3.5 Rotate the transmitter on the sensor (optional)
In integral installations, you can rotate the transmitter on the sensor up to 360º in 45º
increments.
Procedure
1. Using a 4 mm hex key, loosen and remove the clamp securing the transmitter head
in place.
Figure 3-12: Removal of the sensor clamp
2. Gently lift the transmitter straight up, and rotate the transmitter to the desired
position.
You can rotate the transmitter to any of the eight positions, but a stop exists that
will not allow a full 360° rotation.
Figure 3-13: Rotation of the transmitter head
3. Gently lower the transmitter onto the base, confirming that the transmitter is in a
locked position.
Installation Manual Mounting and sensor wiring
MMI-20027478 February 2020
Installation Manual 19