4.10.2 Wire the Frequency Input (externally powered)
Procedure
Wire to the appropriate input terminal and pins.
Figure 4-28: Frequency Input wiring (externally powered)
A
D
B
C
7
8
A. Frequency Input
B. Frequency Input device
C. 1–10 KΩ resistance
D. 3–30 VDC
4.11 Wire Channels A and D for SIS applications
Use this procedure to wire Channels A and D for Safety Instrumented Systems (SIS)
applications.
Procedure
Use one of the following power options:
Channel A active (internal) power and Channel D passive (external) power
Channel A passive (external) power and Channel D passive (external) power
Wiring the channels Installation Manual
February 2020 MMI-20027478
38 Micro Motion 5700 Transmitters with Configurable Outputs