You can rotate the junction box plus or minus 180º to any position.
Figure 3-17: Rotation of the sensor wiring junction box
3. Gently set the junction box into its new position, confirming that the position is
locked.
4. Replace the clamp in its original position and tighten the cap screw. Torque to 28 in
lbf (3.16 N m) to 30 in lbf (3.39 N m).
Figure 3-18: Re-attachment of the clamp
Mounting and sensor wiring Installation Manual
February 2020 MMI-20027478
22 Micro Motion 5700 Transmitters with Configurable Outputs