Prerequisites
The temperature calibration is a two-part procedure: temperature offset calibration and
temperature slope calibration. The two parts must be performed without interruption, in
the order shown. Ensure that you are prepared to complete the process without
interruption. You will need a low-temperature calibration fluid and a high-temperature
calibration fluid. You will not see the effect of the calibration until both the temperature
offset calibration and the temperature slope calibration are complete.
Important
Consult Micro Motion before performing a temperature calibration. Under normal circumstances,
the temperature circuit is stable and should not need an adjustment.
Procedure
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Next
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Calibration Complete
Temperature Slope calibration
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Done
Service Tools >
Maintenance >
Calibration >
Temperature Calibration
Offset
Calibration in Progress
Next
Enter temperature of high-
temperature fluid
Next
Calibration Complete
Service Tools >
Maintenance >
Calibration >
Temperature Calibration
Slope
Calibration in Progress
Finish