EasyManua.ls Logo

Emerson Rosemount 2051G - Switch HART revision mode

Emerson Rosemount 2051G
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
Reference Manual
00809-0300-4410, Rev BA
Configuration
January 2018
Configuration
2.6.2 Configuration tools
Configuring with a Field Communicator
There are two interfaces available with the Field Communicator: Traditional and Dashboard interfaces.
All steps using a Field Communicator will be described using Dashboard interfaces. HART shows the
Device Dashboard interface. As stated in System readiness , it is critical that the latest DD’s are loaded
into the Field Communicator. Visit Emerson.com
or FieldCommGroup.org to download latest DD library.
Field Communicator menu trees and Fast Keys are available in Appendix C: Field Communicator Menu
Trees and Fast Keys.
Figure 2-5. Device Dashboard
Configuring with AMS Device Manager
Full configuration capability with AMS Device Manager requires loading the most current Device
Descriptor (DD) for this device. Download the latest DD at Emerson.com
or FieldCommGroup.org.
Note
All steps using AMS Device Manager will be described using version 11.5.
Configuring with a local operator interface
The LOI requires option code M4 to be ordered. To activate the LOI push either configuration button.
Configuration buttons are located on the LCD display (must remove housing cover to access), or
underneath the top tag of the transmitter. See Tab le 2 -2 for configuration button functionality and
Figure 2-6 for configuration button location. When using the LOI for configuration, several features
require multiple screens for a successful configuration. Data entered will be saved on a screen-by-screen
basis; the LOI will indicate this by flashing “SAVED” on the LCD display each time.
LOI menu trees are available in Appendix D: Local Operator Interface.
SAVE
1. Overview
2. Configure
3. Service Tools
2051 FT 45B
Online

Table of Contents