2.10 Configuring the LCD display
The LCD display configuration command allows customization of the LCD display to suit application
requirements. The LCD display will alternate between the selected items.
In the following instructions, the LCD display can also be configured to display configuration information
during the device startup. Select Review Parameters at Startup to enable or disable this functionality.
Reference Figure 2-2 on page 7 LCD display with LOI for image of LCD display screen.
Field Communicator
AMS Device Manager
1. Right click on the device and select Configure.
2. Select Manual Setup, select the Display tab.
3. Select desired display options and click Send.
LOI
Refer to Figure 2-12 for LCD display configuration using a LOI.
Figure 2-12. Display with LOI
2.11 Detailed transmitter setup
2.11.1 Configuring alarm and saturation levels
In normal operation, the transmitter will drive the output in response to pressure from the lower to
upper saturation points. If the pressure goes outside the sensor limits, or if the output would be beyond
the saturation points, the output will be limited to the associated saturation point.
Pressure Units Sensor Temperature
% of Range mA/Vdc Output
Scaled Variable
From the HOME screen, enter the Fast Key sequence
Device Dashboard Fast Keys
2, 2, 4
DISPLAY
PRESS (on/off)
SCALED (on/off)
TEMP (on/off)
%RANGE (on/off)
ANALOG (on/off)
STRTUP (on/off)
BACK TO MENU
EXIT MENU
VIEW CONFIG
ZERO TRIM
UNITS
RERANGE
LOOP TEST
DISPLAY
DISPLAY
EXTENDED MENU
EXIT MENU