Prerequisites
The O2 cell and heater strut assembly (9, Figure 9-5) and the COe assembly (5, Figure 9-8)
must be removed before you start this procedure.
Procedure
1. Use a straightedge to matchmark alignment of the eductor flat and elbow, as shown
in Figure 9-10.
Eductor alignment matchmarksFigure 9-10:
A. Eductor flat
B. Straight edge
C. Matchmark
2. Unscrew the terminal block mounting (13, Figure 9-8). Move the terminal block
mounting away from the eductor (6).
Maintenance and service
Reference Manual 225