Figure 7-1: Field Communicator Connections - HART
A. Loop connectors
B. Field Communicator
C. Lead set
D. 4-20 mA terminal block in electronics housing
E. O
2
4-20 mA signal line
F. Analog output device
G. Field Communicator rear panel
7.2.2 
Connecting to a FOUNDATION
Fieldbus segment
Connect the Field Communicator with the supplied lead set in parallel with the
device to a Fieldbus segment, Figure 7-2. The Field Communicator Fieldbus
connections are polarity sensitive; an error message displays if the device is
connected incorrectly.
WARNING
EXPLOSION
Explosions can result in death or serious injury.
Do not make connections to the Field Communicator's serial port, digital signal
line, or NiCad recharger jack in an explosive atmosphere.
Field Communicator Manual
December 2022 00809-0500-4880
106 Emerson.com/Rosemount