1.3 Specifications
Note
All static performance characteristics are with operating variables constant.
Specifications subject to change without notice.
1.3.1 Net O
2
range
0-1 percent to 0-40 percent O
2
, fully field selectable
1.3.2 Combustibles
0-1000 ppm to 0-5 percent, fully field selectable
1.3.3 Accuracy
Oxygen
±0.75 percent of reading or 0.5 percent O
2
(whichever is greater)
Combustibles
±50 ppm
1.3.4
System response to test gas
Oxygen
10 seconds T90
Combustibles
25 seconds T90
1.3.5
Temperature limits
Process
32 to 2600 °F (0 to 1427 °C)
Sensors housing
-40 to 212 °F (-40 to 100 °C) ambient
Electronics housing
-40 to 149 °F (-40 to 65 °C), ambient
Description and specifications Manual
December 2022 00809-0500-4880
22 Emerson.com/Rosemount