9. At the prompt, input a time interval (in hours) at which an automatic COe
calibration will occur and press ENTER.
To disable automatic calibration for O
2
and COe, enter 0 for both CalIntrvl
parameters.
6.3  Operator-initiated autocalibration
An operator can initiate an automatic calibration at any time, as long as the
transmitter is equipped with calibration solenoids.
6.3.1  Autocalibrate using the optional local operator
interface (LOI)
To initiate a calibration using the LOI, perform the following steps on the LOI
menu tree. Refer to Figure 4-4 for the LOI menu tree.
Procedure
1. From the Calibration menu, press Right to select the Cal Control menu.
2. Select Start Cal-O2, Start Cal COe, or Start Cal Both to start the calibration.
Select Cal Verify to access the Calibration window.
3. At the prompt, press Right to initiate automatic calibration.
6.3.2 
Autocalibrate using HART
®
To initiate an automatic calibration using HART/AMS, perform the following steps
on the HART menu tree. Refer to Figure 5-3 for the HART menu tree.
Procedure
1. From the DEVICE SETUP menu, select DIAG/SERVICE.
2. From the DIAG/SERVICE menu, select CALIBRATION.
3. From the CALIBRATION menu, select PERFORM CAL.
4. From the PERFORM CAL menu, select CAL METHODS.
5. From the CAL METHODS menu, select the type of calibration desired: O2
Calibration, COe Calibration, or O2 and COe Calibration.
6.4  Manual calibration
If a transmitter is not equipped with calibration solenoids, an operator must
calibrate by following prompts from the transmitter.
Refer to Calibrate manually using the optional local operator interface (LOI),
Calibrate O
2
manually using the Field Communicator: HART
®
, or Calibrate COe
manually using the Field Communicator: HART
®
.
Manual
Calibration
00809-0500-4880 December 2022
Rosemount OCX8800A 95