8.7  Transducer block enumerations
8.7.1  Calibration states
During a running calibration procedure, the states below reflect the current step
that the calibration is running in.
Refer to Table 8-4.
Table 8-4: Calibration State: Values
Index CAL_STATE description Operator acknowledgement
required to continue?
0 Idle Yes, calibration can also be
initiated from internally generated
events.
1 Apply O
2
low gas Yes, if parameter Solenoids
Present is 0.
2 Flow O
2
low gas No
3 Read O
2
low gas No
4 Apply O
2
high gas Yes, if parameter Solenoids
Present is 0.
5 Flow O
2
high gas No
6 Read O
2
high gas No
7 Apply combustibles low gas Yes, if parameter Solenoids
Present is 0.
8 Flow combustibles low gas No
9 Read combistibles low gas No
10 Apply combustibles high gas Yes, if parameter Solenoids
Present is 0..
11 Flow combustibles high gas No
12 Read combustibles high gas No
13 Stop gas Yes, if parameter Solenoids
Present is 0.
14 Purge No
15 Abort Yes, if parameter Solenoids
Present is 0.